Abstract
In this work, two dissolved oxygen sensors containing different electrolytes one of 1M KCL and the other 1 M KOH are used to measure dissolved oxygen in water at 290C in Makurdi town. The current was measured with digital meter to four decimal places. The current is directly proportional to the partial pressure of dissolved oxygen. This was done to observe the behaviour of such a sensor in a tropical and humid town like Makurdi. The sensor with KOH had a greater sensitivity detecting currents in the order of milliamperes without amplification compared to the sensor with KCl which indicated unamplified current in order of microampere. The histogram and data shows that the sensor with KOH has less scatter and also less deviation from the mean and hence recommended for use although care must be taken in design considering the corrosive nature of KOH in its tendency to cause more wears. In terms of instrumentation, the KOH sensor has a greater precision and resolution under the prevailing circumstance. This work indicates that these sensors are suitable for dissolved oxygen sensors and instruments calibrated to one or two decimal places which is very valuable to monitor oxygen in fish ponds.